Currently, a 3 x 3 BGO detector is
incorporated into the design of the PELAN. However, the high γ-ray efficiency of this
detector has a drawback in that it detects γ-rays from the environment
surrounding the object under interrogation. These γ-rays from the environment
are essentially noise that must be filtered from the signal from the
object under interrogation. Since there is no practical way to
focus the neutrons the signal-to-noise ratio (SNR) of the
detector must be modified. We have tried
several approaches to solve this problem.
The traditional way of removing the noise
from the ground or from the neutron generator is to place a very
heavy, dense material such as lead between the objects creating the
noise and the detector. Unfortunately, this has two drawbacks: 1)
for a portable system, this increases the weight, and 2) these
materials contribute to the noise.
Another way of removing the noise from the
spectrum is to use an active filtering device. For example,
Compton-suppression systems are often used in nuclear physics to
eliminate the continuum caused by Compton scattering within the
detector.
Carbon and oxygen measurements are most
likely to be affected by the noise from the environment and the
neutron generator.Thus, an active filter must be optimized for
γ-rays with energies
between 4 and 6 MeV. In this energy range, the two dominant photon
interactions with matter are pair production and scattering
(primarily Compton scattering).
Figure 1 is a schematic drawing showing
γ-rays entering the
detector. On the left-hand side (A), three γ-rays are entering PELAN’s
detector. The top-most γ-ray
is scattered and there is an appreciable probability that the
Compton scattered γ-ray
will leave the detector. The bottom-most γ-ray undergoes pair
production with a 511 keV γ-ray
escaping the detector. Only the γ-ray
that is perpendicular to the front surface of the detector deposits
its entire energy within the detector.
Figure 1. Schematic representation of high energy
γ-rays entering the PELAN
detector (A). The PELAN
detector surrounded by a veto detector (B). See text for details.
|
On the right-hand side of Figure 1, the same detector is now
surrounded by another detector. The two γ-rays that deposited only
a fraction of their energy have a high probability of being absorbed
in this second detector. Data are collected in the PELAN detector
only when there is no corresponding signal from this second detector
(anti-coincidence mode). Since this second detector vetoes data in
the PELAN detector, we are describing it as a “veto detector”.
Figure 2 shows a veto detector
designed for the PELAN project. The PELAN detector is a 7.6 cm
x 7.6 cm (3”x3”) BGO detector. The veto detector is also
composed of BGO with a thickness of 2.5 cm (1”) with an inner
diameter just over 7.6 cm
Figure 2. The PELAN detector surrounded by the veto detector.
|
The veto detector is comprised of four optically isolated pieces of
BGO. Each piece has two photomultiplier tubes in order to
efficiently gather the light from the scintillations. In
practice, the signals from the pair of the photomultipliers from
each piece are tied together. There is no attempt to gather
position information from the veto detector.
An anti-coincidence circuit for the veto
detector has also been designed. The timing resolution between
PELAN’s detector and the veto detector is approximately 10 ns when
using a 22Na γ-ray
source.
The veto detector is not fully annular and
only completes three-quarters of the circumference. Tests completed
during the design phase of the veto detector construction indicated
that a complete annulus was not required. This is due to the
shielding intrinsic to PELAN which is sufficient to
limit the γ-ray noise
emanating from the neutron generator.
Figure 3 shows two
γ-ray spectra: the dashed
line spectrum is taken without the veto detector, and the solid line
spectrum is taken with the veto detector surrounding the PELAN
detector. The shielding effect of the veto detector is apparent,
since the solid line spectrum is lower than the dashed line one.
In the region around 5.1 MeV, the second escape peak of the 6.1 MeV
γ-ray is totally absent
from the spectrum taken with the veto detector.
Figure 3. The spectrum from the PELAN detector in anti-coincidence mode with veto
detector (solid line) and a spectrum taken with the PELAN detector alone.
Both spectra were taken with the detector near the ground.
|
The
veto detector does indeed reduce the “noise” generated by the
background. However, the change in the SNR was equivalent to the
SNR produced by surrounding the BGO with a 1” thick piece of Pb. We
could not justify the expense and the complexity in utilizing the
veto detector.
A third way is to change the geometry of
the detector. The gamma rays that PELAN analyzes range in energy
from 1.0 MeV to 8 MeV. These are highly penetrating and require a
long free mean path within the BGO for full energy deposition in the
crystal. If the geometry is changed so that the cross-sectional
area of the detector is smaller than the length of the detector,
then the higher energy gamma rays will deposit preferentially their
full energy along the long axis of the detector (see Figure 4). A
large detector can then be created having a close-packed array of
these detectors. To suppress scattering, any event in which two or
more detectors are in coincidence is vetoed out. In this manner,
the solid angle is maintained and gamma rays in the direction of the
long axis are preferentially selected. We refer to this design as a
“segmented” detector.
Figure 4. Segmented detector.
|
However,
there are several technical problems associated with a segmented
detector. First, the resolution of the detector decreases due
to the fact that smaller diameter photo-multiplier tubes have a
poorer light collection efficiency. Because of this limitation, the
segments could not have a diameter much smaller than 2.5 cm (1”)
with a 7.6 cm (3”) length. Second, the detectors must be in
close proximity to increase the efficiency of the anti-coincidence
circuit. Finally, compact PC-based electronics must be
designed.
A proof of principle test was performed utilizing 4 NaI detectors
which matched the desired shape. They were connected to an
anti-coincidence circuit. The analysis of the effectiveness of this detector is
still ongoing.
For PELAN, we are still seeking an improved
detection system. We desire improvements in directionality which we
believe will change the SNR ratio and eliminate the problems caused
by clutter.
Acknowledgements:
This work was
supported in part by the Department of Defense under contracts
DAAD07-98-C-0116, DAAD05-98-C-022 and DACA72-01-C-0017
For more detailed
information, please see the
Optimizing the Signal to Noise Ratio for the PELAN System
Publications
portion of this web site.